Journal of the Faculty of Science, Hokkaido University. Series 4, Geology and mineralogy - Vol.XXII, No.3

Major Element Analysis of Rock Samples by X-ray Fluorescence Spectrometry, using Scandium Anode Tube

Tsuchiya, Nobutaka;Shibata, Tomoyuki;Koide, Yoshiyuki;Owada, Masaaki;Takazawa, Eiichi;Goto, Yoshihiko;Choi, Jai Ho;Terada, Shoichi;Hariya, Yu

Permalink :  http://hdl.handle.net/2115/36761

Abstract

An X-ray fluorescence spectrometer (Philips PW1404) equipped with a scandium anode tube was tested for major element analysis of rock samples. Rock powder, five times flux (Li2B4O7), and 20 to 30mg of LiBr・H2O were mixed and fused to make a glass bead. The calibration lines were determined from 13 rock standards of GSJ and USGS and 14 synthetic standards by using a software X44 for the PW1404. The synthetic standards were useful in estimating of the matrix correction factors, and adequate calibration lines were provided for a wide range of composition. The analytical errors of the calibration lines were enough small for the purpose of petrological investigations. Using an automatic sample changer PW1500 and a software X44 for XRF, more than 80 rock samples can be automatically analyzed in a day. This method is suitable for the routine work of major element analysis of rock samples.


FULL TEXT :  PDF