Title: Aberration-corrected multipole Wien filter for energy-filtered x-ray photoemission electron microscopy
Authors: Niimi, Hironobu Browse this author
Chun, Wang-Jae Browse this author
Suzuki, Shushi Browse this author
Asakura, Kiyotaka Browse this author
Kato, Makoto Browse this author
Issue Date: Jun-2007
Publisher: American Institute of Physics
Journal Title: Review of Scientific Instruments
Volume: 78
Issue: 6
Start Page: 063710
Publisher DOI: 10.1063/1.2749449
PMID: 17614619
Abstract: The aberration of a multipole Wien filter for energy-filtered x-ray photoemission electron microscopy was analyzed and the optimized Fourier components of the electric and magnetic fields for the third-order aperture aberration corrections were obtained. It was found that the third-order aperture aberration correction requires 12 electrodes and magnetic poles. ©2007 American Institute of Physics