A Proton Beam Therapy System Dedicated to Spot-Scanning Increases Accuracy with Moving Tumors by Real-Time Imaging and Gating and Reduces Equipment Size

This item is licensed under:Creative Commons Attribution 3.0 Unported

Files in This Item:
PLoS One_9(4)_e94971.pdf897.8 kBPDFView/Open
Please use this identifier to cite or link to this item:http://hdl.handle.net/2115/56313

Title: A Proton Beam Therapy System Dedicated to Spot-Scanning Increases Accuracy with Moving Tumors by Real-Time Imaging and Gating and Reduces Equipment Size
Authors: Shimizu, Shinichi Browse this author →KAKEN DB
Miyamoto, Naoki Browse this author →KAKEN DB
Matsuura, Taeko Browse this author →KAKEN DB
Fujii, Yusuke Browse this author
Umezawa, Masumi Browse this author
Umegaki, Kikuo Browse this author →KAKEN DB
Hiramoto, Kazuo Browse this author
Shirato, Hiroki Browse this author →KAKEN DB
Issue Date: 18-Apr-2014
Publisher: The Public Library of Science
Journal Title: PLoS ONE
Volume: 9
Issue: 4
Start Page: e94971
Publisher DOI: 10.1371/journal.pone.0094971
PMID: 24747601
Abstract: PURPOSE: A proton beam therapy (PBT) system has been designed which dedicates to spot-scanning and has a gating function employing the fluoroscopy-based real-time-imaging of internal fiducial markers near tumors. The dose distribution and treatment time of the newly designed real-time-image gated, spot-scanning proton beam therapy (RGPT) were compared with free-breathing spot-scanning proton beam therapy (FBPT) in a simulation. MATERIALS AND METHODS: In-house simulation tools and treatment planning system VQA (Hitachi, Ltd., Japan) were used for estimating the dose distribution and treatment time. Simulations were performed for 48 motion parameters (including 8 respiratory patterns and 6 initial breathing timings) on CT data from two patients, A and B, with hepatocellular carcinoma and with clinical target volumes 14.6 cc and 63.1 cc. The respiratory patterns were derived from the actual trajectory of internal fiducial markers taken in X-ray real-time tumor-tracking radiotherapy (RTRT). RESULTS: With FBPT, 9/48 motion parameters achieved the criteria of successful delivery for patient A and 0/48 for B. With RGPT 48/48 and 42/48 achieved the criteria. Compared with FBPT, the mean liver dose was smaller with RGPT with statistical significance (p<0.001); it decreased from 27% to 13% and 28% to 23% of the prescribed doses for patients A and B, respectively. The relative lengthening of treatment time to administer 3 Gy (RBE) was estimated to be 1.22 (RGPT/FBPT: 138 s/113 s) and 1.72 (207 s/120 s) for patients A and B, respectively. CONCLUSIONS: This simulation study demonstrated that the RGPT was able to improve the dose distribution markedly for moving tumors without very large treatment time extension. The proton beam therapy system dedicated to spot-scanning with a gating function for real-time imaging increases accuracy with moving tumors and reduces the physical size, and subsequently the cost of the equipment as well as of the building housing the equipment.
Rights: http://creativecommons.org/licenses/by/3.0/
Type: article
URI: http://hdl.handle.net/2115/56313
Appears in Collections:国際連携研究教育局 : GI-CoRE (Global Institution for Collaborative Research and Education : GI-CoRE) > 雑誌発表論文等 (Peer-reviewed Journal Articles, etc)
医学院・医学研究院 (Graduate School of Medicine / Faculty of Medicine) > 雑誌発表論文等 (Peer-reviewed Journal Articles, etc)

Submitter: 清水 伸一

Export metadata:

OAI-PMH ( junii2 , jpcoar_1.0 )

MathJax is now OFF:


 

 - Hokkaido University