HUSCAP logo Hokkaido Univ. logo

Hokkaido University Collection of Scholarly and Academic Papers >
Office of Health and Safety >
Peer-reviewed Journal Articles, etc >

Optimization of helical acquisition parameters to preserve uniformity of mouse whole body using multipinhole collimator in single-photon emission computed tomography

This item is licensed under:Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International

Files in This Item:
1-s2.0-S2211379716301437-main.pdf642.59 kBPDFView/Open
Please use this identifier to cite or link to this item:http://hdl.handle.net/2115/64293

Title: Optimization of helical acquisition parameters to preserve uniformity of mouse whole body using multipinhole collimator in single-photon emission computed tomography
Authors: Ukon, Naoyuki Browse this author
Kubo, Naoki Browse this author →KAKEN DB
Ishikawa, Masayori Browse this author →KAKEN DB
Zhao, Songji Browse this author
Tamaki, Nagara Browse this author →KAKEN DB
Kuge, Yuji Browse this author →KAKEN DB
Keywords: Helical acquisition
Multipinhole collimator
Computed tomography
SPECT
Issue Date: 20-Sep-2016
Publisher: Elsevier
Journal Title: Results in physics
Volume: 6
Start Page: 659
End Page: 663
Publisher DOI: 10.1016/j.rinp.2016.09.010
Abstract: Focusing on whole-body uniformity in small-animal single-photon emission computed tomography (SPECT), we examined the optimal helical acquisition parameters using five-pinhole collimators for mouse imaging. SPECT images of an 80-mm-long cylindrical phantom with 99mTc solution were acquired using an Inveon multimodality imaging platform. The bed travels used in this study were 0, 30, 60, 90 and 120 mm, and the numbers of revolutions traversed during the SPECT scan were 1.0, 2.0, 3.0, 4.0, 5.0 and 7.0, respectively. Artifacts that degrade uniformity in reconstructed images were conspicuous when the bed travel was smaller than the object length. Regarding the distal-to-center ratio (DCR) of SPECT values in the object's axial direction, the DCR nearest to the ideal ratio of 1.00 was 1.02 in the optimal uniformity with 4.0 revolutions and a bed travel of 120 mm. Moreover, the helical acquisition using these parameters suppressed the formation of artifacts. We proposed the optimal parameters in whole-body helical SPECT; the bed travel was sufficiently larger than the object length; the 4.0 or more revolutions were required for a pitch of approximately 30 mm/revolution. The optimal acquisition parameters in SPECT to preserve uniformity would contribute to the accurate quantification of whole-body biodistribution. (C) 2016 The Authors. Published by Elsevier B.V. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
Rights: http://creativecommons.org/licenses/by-nc-nd/4.0/
Type: article
URI: http://hdl.handle.net/2115/64293
Appears in Collections:安全衛生本部 (Office of Health and Safety) > 雑誌発表論文等 (Peer-reviewed Journal Articles, etc)
国際連携研究教育局 : GI-CoRE (Global Institution for Collaborative Research and Education : GI-CoRE) > 雑誌発表論文等 (Peer-reviewed Journal Articles, etc)

Submitter: 久保 直樹