HUSCAP logo Hokkaido Univ. logo

Hokkaido University Collection of Scholarly and Academic Papers >
Graduate School of Engineering / Faculty of Engineering >
Peer-reviewed Journal Articles, etc >

Determination of Cadmium in Brown Rice Samples by Fluorescence Spectroscopy Using a Fluoroionophore after Purification of Cadmium by Anion Exchange Resin

This item is licensed under:Creative Commons Attribution 4.0 International

Files in This Item:
sensors-17-02291-v2.pdf762.48 kBPDFView/Open
Please use this identifier to cite or link to this item:http://hdl.handle.net/2115/68047

Title: Determination of Cadmium in Brown Rice Samples by Fluorescence Spectroscopy Using a Fluoroionophore after Purification of Cadmium by Anion Exchange Resin
Authors: Hafuka, Akira Browse this author
Takitani, Akiyoshi Browse this author
Suzuki, Hiroko Browse this author
Iwabuchi, Takuya Browse this author
Takahashi, Masahiro Browse this author
Okabe, Satoshi Browse this author →KAKEN DB
Satoh, Hisashi Browse this author →KAKEN DB
Keywords: cadmium
rice
simple analytical method
fluorescence spectroscopy
pretreatment
Issue Date: 9-Oct-2017
Publisher: MDPI
Journal Title: Sensors
Volume: 17
Issue: 10
Start Page: 2291
Publisher DOI: 10.3390/s17102291
Abstract: Simple analytical methods are needed for determining the cadmium (Cd) content of brown rice samples. In the present study, we developed a new analytical procedure consisting of the digestion of rice using HCl, Cd purification using anion exchange resin, and then determining the Cd content using fluorescence spectroscopy. Digestion with 0.1 M HCl for 10 min at room temperature was sufficient to extract Cd from the ground rice samples. The Cd in the extract was successfully purified in preference to other metals using Dowex 1X8 chloride form resin. Low concentrations of Cd in the eluate could be determined using fluorescence spectroscopy with a fluoroionophore. Overall, the actual limit of quantification value for the Cd content in rice was about 0.1 mg-Cd/kg-rice, which was sufficiently low compared with the regulatory value (0.4 mg-Cd/kg-rice) given by the Codex Alimentarius Commission. We analyzed authentic brown rice samples using our new analytical procedure and the results agreed well with those determined using inductively coupled plasma optical emission spectrometry (ICP-OES). Since the fluoroionophore recognized Zn2+ and Hg2+ as well as Cd2+, a sample containing high concentration of Zn2+ or Hg2+ might cause a false positive result.
Rights: https://creativecommons.org/licenses/by/4.0/
Type: article
URI: http://hdl.handle.net/2115/68047
Appears in Collections:工学院・工学研究院 (Graduate School of Engineering / Faculty of Engineering) > 雑誌発表論文等 (Peer-reviewed Journal Articles, etc)

Submitter: 佐藤 久

Export metadata: